OBJECTIVES: Penile prosthesis implantation is the gold standard treatment for erectile dysfunction in patients whose first and secondline treatment options gave unsuccessful results. Mechanical failure with a malleable penile prosthesis is very rare. To our knowledge in the literature only one case was reported a bilateral complete rod fracture of malleable penile prosthesis.
CASE PRESENTATION: This is the second case in literature reporting a bilateral complete rod fracture of malleable penile prosthesis. We removed the previous pieces and implanted a new malleable device during the same session.
CONCLUSIONS: We learned very important lessons with this case; mechanical failure of the malleable penile prosthesis can be seen as a late complication of the surgery. Rod fracture should be kept in mind in the patients who are not comfortable during sexual intercourse.
GİRİŞ: Penis protezi implantasyonu, birinci ve ikinci basamak tedavi seçenekleri başarısız olan hastalarda erektil disfonksiyon için altın standart tedavidir. Malleabl bir penil protez ile mekanik arıza çok nadirdir. Literatürde bilgilerimize göre sadece bir olguda çift taraflı malleable penil protezin komplet fraktürü bildirilmiştir.
OLGU: Bu, literatürde malleabl penil protezin iki taraflı komplet fraktürü bildiren ikinci vakadır. Aynı seansta önceki parçaları çıkardık ve yeni bir malleabl cihaz yerleştirdik.
SONUÇ: Bu vakadan çok önemli dersler çıkardık; malleabl penil protez fraktürü ameliyatın geç bir komplikasyonu olarak görülebilir. Cinsel ilişki sırasında rahat olmayan hastalarda protez fraktürü akılda tutulmalıdır.